The Bruker PharmaScan MRI system is equipped with an horizontal 7 Tesla Magnet, with a bore of 16 cm diameter. The Pharmascan system is based on Avance NEO electronics and works with Bruker BGA 9S HP gradients yielding a maximum strength of 760 mT / m. The scanner has 4 channels for transmission and reception of signals and can detect both protons and X nuclei such as 13C, 19F, 23Na. The samples can be located inside the magnet bore in a very precise and reproducible way with the ManPac system. The high magnetic field of the PharmaScan increases the sensitivity of MRI studies and allows the investigation of millimetric structures with a suitable resolution. Moreover, it also improves the sensitivity in other advanced applications depending on the Chemical Shift, such as MRS, and from local Susceptibility, such as the fMRI with BOLD (Blood Oxygenation Level Dependent) contrast.